A. Dirt or dust on compression paddle
B. Ghost image
C. Readout failure
D. Dead pixels
E. Gridlines
50
 What is the artifact present on the following mediolateral oblique (MLO) image? The second image denotes a part of the MLO view magnified.

A. Motion
B. Gridlines
C. Deodorant
D. Filtration artifact
51
 The universal amount of intravenous gadolinium used for contrast enhancement in breast MR imaging is:
A. 0.1 mmol/kg
B. 0.2 mmol/kg
C. 0.3 mmol/kg
D. 0.4 mmol/kg
E. 0.5 mmol/kg
52
 In order to ensure the quality of the mammographic images, the posterior nipple line on MLO and CC projections should be within
A. 0.5 cm
B. 1.0 cm
C. 1.5 cm
D. 2.0 cm
E. It should be equal.
53a
 Which one of the following artifacts is present on the T1-weighted non–fat-saturated localizer image?

A. Chemical shift
B. Phase wrap/aliasing
C. Metallic susceptibility
D. Patient motion
53b
 What can reduce phase wrap/aliasing artifact on breast MRI?
A. Enlarging the field of view
B. Reducing patient motion
C. Shimming the magnet frequently
D. Increasing the bandwidth
E. Check for a leak in the radiofrequency (RF) shield
54
 Which feature of digital breast tomosynthesis allows it to decrease the effect of overlapping breast tissue?
A. Higher radiation dose than in mammography
B. Reconstruction of the projections into the mediolateral oblique (MLO) and craniocaudal (CC) views
C. Digital acquisition technique
D. Multiple exposures of the breast at different angles
E. Increased breast compression
55
 The above mediolateral oblique (MLO) image taken during a screening mammogram examination demonstrates which type of digital mammogram artifact?

A. Detector interface line
B. Ghost image
C. Readout failure
D. Dead pixels
E. Gridlines
ANSWERS AND EXPLANATIONS
1

Answer C.
References:
www.fda.gov/cdrh/mammography
Ikeda D.
Breast Imaging: The Requisites
. 2nd ed. St. Louis, MO: Elsevier Mosby; 2011:16.

2

A. Answer A.
BI-RADS 2—Dermal calcifications
B. Answer A.
BI-RADS 2—“Popcorn-like” calcification
C. Answer A.
BI-RADS 2—Secretory calcifications
D. Answer A.
BI-RADS 2—Round calcification
E. Answer A.
BI-RADS 2—Lucent center calcification (oil cyst)
F. Answer B.
BI-RADS 4—Linear branching calcifications
G. Answer B.
BI-RADS 4—Fine pleomorphic calcifications
H. Answer B.
BI-RADS 2—Milk of calcium
Reference: American College of Radiology (ACR). BI-RADS-Mammography.
The ACR Breast Imaging Reporting and Data System (BI-RADS)
. Reston, VA: American College of Radiology; 2003:61–107.
3

Answer B.
Reference: Linver MN, Osuch JR, Brenner RJ, et al. The mammography audit: A Primer for the Mammography Quality Standards Act (MQSA).
AJR Am J Roentgenol
1995;165:19–25.
4

Answer D.
 Sensitivity is the probability of detecting a cancer when a cancer exists or the number of cancers diagnosed after being identified at mammography in a population within 1 year of the imaging examination, divided by all cancers present in that population in the same time period.

Sensitivity = TP/(TP + FN); TP = True positive; FN = False negative

References: American College of Radiology (ACR). BI-RADS—Mammography.
The ACR Breast Imaging Reporting and Data System (BI-RADS)
. Reston, VA: American College of Radiology; 2003:231.
Linver MN, Osuch JR, Brenner RJ, et al. The mammography audit: A Primer for the Mammography Quality Standards Act (MQSA).
AJR Am J Roentgenol
1995;165:19–25.
5

Answer C.
 The appearance of the pectoral muscle is useful in assessing position on MLO views. The breast should be pulled up and out, the inframammary fold should be open on MLO views and neutral on craniocaudal views, and a small amount of the upper abdomen should be visible on MLO views.
Reference: Cardenosa G.
Breast Imaging Companion
. 3rd ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2008:79–84.
6

Answer A.
 A breast cancer diagnosed within a year of a negative screening mammogram is considered a false negative.
Reference: BI-RADS-Mammography.
The ACR Breast Reporting and Data System (BI-RADS)
. Reston, VA: American College of Radiology.
Linver MN, Osuch JR, Brenner RJ, et al. The mammography audit: A Primer for the Mammography Quality Standards Act (MQSA).
AJR Am J Roentgenol
1995;165:19–25.
7

Answer C.

Quality Control (QC) Test Schedule for Film-Screen Mammography
TEST
PERFORMED
Darkroom cleanliness
Daily
Processor QC
Daily
Screen cleanliness
Weekly
Phantom images
Weekly
Viewbox cleanliness and viewing checklist
Quarterly
Repeat analysis
Quarterly
Fixer retention
Semiannually
Darkroom fog
Semiannually
Screen–film contact
Semiannually
Compression
Semiannually
BOOK: Breast Imaging: A Core Review
6.34Mb size Format: txt, pdf, ePub
ads

Other books

Pearls by Colin Falconer
Promissory Payback by Laurel Dewey
Seducing a Wolf: Moonbound Series, Book Five by Krystal Shannan, Camryn Rhys
Whipped) by Karpov Kinrade
Oxfordshire Folktales by Kevan Manwaring
Unlikely Praise by Carla Rossi
Too Consumed by Skyla Madi
El equipaje del rey José by Benito Pérez Galdós